Peristaltic pump tubing and sample line supplies

Dodávateľ: AGILENT HEWLETT PACKARD
G3280-40103 G8400-40113
HEWLG3280-40103EA 191.87 EUR
HEWLG3280-40103 HEWLG8400-40113
Peristaltic pump tubing and sample line supplies
Chromatografické potřeby Chromatography Spare Parts
The peristaltic pump tubing has to be washed prior to use. The common way is to run about 5% HNO₃ (v/v) solution.

  • Long lifetime
  • Resistant to most chemicals
  • Not suitable for sample introduction - impurity level is high

If customers use HCl, Sn is eluted from the Tygon tube. In this case, 3% HCl (v/v) should be pumped at 0,1 rps for less than an hour by the new peristaltic pump tube.

Pozor: Some chemicals attack the peristaltic pump tube. This might cause leakage in the sample introduction and drain lines and corrosion of the sample introduction system.